Set–up Error for Prostate Cancer Patients in Intensity–modulated Radiation Therapy with Pelvis Mask

Authors

  • Wimonmart Tong–Ngarm Sakon Nakhon Hospital
  • Sarayut Kornsopa Sakon Nakhon Hospital
  • Phattanapong Saenchon Sakon Nakhon Hospital
  • Kiattisak Promsensa Sakon Nakhon Hospital

Keywords:

Setup error, Intensity–modulated radiotherapy, Prostate cancer

Abstract

             The research aimed to study the set–up error in positioning for intensity–modulated radiation therapy (IMRT) with pelvic mask in prostate cancer patients to obtain an appropriate value of CTV to PTV margin for radiation therapy in Sakon Nakhon hospital. The set–up error values from patient positioning with pelvic mask for IMRT in prostate cancer patients were retrospectively recorded by comparing anteroposterior (AP) and lateral (LAT) images from an electronic portal imaging device (EPID) and digital reconstructed radiographs (DRRs). The recorded values were used to calculate CTV to PTV margin using Van Herk equation. The samples were 574 EPID and 82 DRRs images obtained from 41 prostate cancer patients.

               The results of the range of set–up error values in the craniocaudal, lateral, and anteroposterior axis were –0.6 to 0.9 cm, –0.6 to 0.6 cm, and –0.7 to 0.4 cm, respectively. The average set–up error and standard deviation in the craniocaudal, lateral, and anteroposterior axis were 0.08 ± 0.26, 0.00 ± 0.17, and –0.03 ± 0.16 cm, respectively. For all three axes, the median and quartile range was 0.0 ± 0.2 cm. The calculated CTV to PTV margins of the craniocaudal, lateral, and anteroposterior axis was 0.47, 0.24, and 0.21 cm, respectively. In conclusion, the set–up errors in positioning for IMRT with pelvic mask in prostate cancer patients were –0.7 to 0.9 cm in all three axes, and the CTV to PTV margins were within the recommended range from ICRU No.62.                 

Author Biographies

Wimonmart Tong–Ngarm, Sakon Nakhon Hospital

Radiological Technologist (Professional Level), Radiotherapy unit, Department of Radiology, Sakon
Nakhon Hospital

Sarayut Kornsopa, Sakon Nakhon Hospital

Radiological Technologist (Professional Level), Radiotherapy unit, Department of Radiology, Sakon
Nakhon Hospital

Phattanapong Saenchon, Sakon Nakhon Hospital

Radiological Technologist (Senior Professional Level), Radiotherapy unit, Department of Radiology,
Sakon Nakhon Hospital

Kiattisak Promsensa, Sakon Nakhon Hospital

Medical doctor (Professional Level), Radiotherapy unit, Department of Radiology, Sakon Nakhon
Hospital

References

กรมการแพทย์กระทรวงสาธารณสุข. แผนการป้องกันและควบคุมโรคมะเร็งแห่งชาติ National Cancer Control Programme (พ.ศ. 2561–2565) [อินเทอร์เน็ต]. 2565 [เข้าถึงเมื่อ 1 สิงหาคม 2565]. เข้าถึงได้จาก: https:// www.nci.go.th/th/File_download/D_index/แผนการป้องกันและควบคุมโรคมะเร็ง

แห่งชาติ.pdf

แผนกตรวจสุขภาพโรงพยาบาลเปาโล พระประแดง. 5 อันดับ สถิติมะเร็งในคนไทย. [อินเตอร์เน็ต]. 2561 [เข้าถึงเมื่อ 1 สิงหาคม 2565]. เข้าถึงได้จาก: https://www.paolohospital.com/th–TH/phrapradaeng/Article/Details/บทความ–โรคมะเร็ง/5–อันดับ–สถิติมะเร็งในคนไทย–ที่ควรดูแลตนเองให้ห่างไกล

หน่วยมะเร็งวิทยา ภาควิชาอายุรศาสตร์ คณะแพทยศาสตร์โรงพยาบาลรามาธิบดีมหาวิทยาลัยมหิดล. มะเร็งต่อมลูกหมาก โรคร้ายของสุขภาพเพศชาย. [อินเตอร์เน็ต]. 2564 [เข้าถึงเมื่อ 10 พฤษภาคม 2565]. เข้าถึงได้จาก: https://www.rama.mahidol.ac.th/cancer_center/sites/default/files/public/pdf/Booklet/

Booklet–Prostate–Cancer.pdf

Saad A, Goldstein J, Lawrence Y, Spieler B, Leibowitz–Amit R, Berger R, et al. Classifying high–risk versus very high–risk prostate cancer: Is it relevant to outcomes of conformal radiotherapy and androgen deprivation?. Radiat Oncol 2017;12:1–8.

Arias–Stella J, Shah A, Montoya–Cerrillo D, Williamson S, Gupta N. Prostate Biopsy and Radical Prostatectomy Gleason Score Correlation in Heterogenous Tumors: Proposal for a Compoite Gleason Score. AJSP 2015;39:1213–8.

American Joint Committee on Cancer. AJCC Cancer Staging Manual. 6th ed. New york: Lippincott Raven; 2002.

Bhardwaj AK, Kehwar TS, Chakarvarti SK, Oinam AS, Sharma SC. 3–Dimensional conformal radiotherapy versus intensity–modulated radiotherapy for localized prostate cancer: Dosimetric and radiobiologic analysis. Int J Radiat Res 2007;5(1):1–8

ศิวลี สุริยาปี, ทวีป แสงแห่งธรรม, พันทิวา อุณห์ศิริ. ฟิสิกส์ทางรังสีรักษา (Physics of Radiotherapy). กรุงเทพฯ: Ideol Digital; 2563.

International Commission on Radiation Units and Measurements. ICRU Report 62: Prescribing recording and Reporting Photon Beam Therapy (Supplement to ICRU Report 50). USA: ICRU; 1999.

Malone S, Szanto J, Perry G, Gerig L, Manion S, Dahrouge S, Crook J. A prospective comparison of three systems of patient immobilization for prostate radiotherapy. Int J Radiat Oncol Biol Phys 2000;48(3):657–65.

Virkar M, Kumar N A, Chadha P, Rodrigues RJ. Vacuum and thermoplastic mould–base immobilization systems used in patient undergoing pelvis radiation therapy: A comparative study. IJCBR 2020;6(1):8–10.

Yamane, Taro. Statistics: An introductory analysis, 2nd Edition, New York: Harper and Row; 1967.

White P, Yee CK, Shan LC, Chung LW, Man NH, Cheung YS. A comparison of two systems of patient immobilization for prostate radiotherapy. Radiat Oncol 2014;9:29.

Herk MV, Remeijer P, Rasch C, Lebesque JV. The probability of correct target dosage: dose–population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phy 2000;47(4):1121–35.

วิมลมาศ ทองงาม, สุมาลี ยับสันเทียะ : การเปรียบเทียบความคลาดเคลื่อนของการจัดท่าผู้ป่วยมะเร็งต่อมลูกหมากด้วยอุปกรณ์ยึดตรึงสามชนิดในเทคนิคการฉายแบบปรับความเข้ม. J Thai Assos Radiat Oncol 2021;2:102–15.

Downloads

Published

2024-06-25

How to Cite

1.
Tong–Ngarm W, Kornsopa S, Saenchon P, Promsensa K. Set–up Error for Prostate Cancer Patients in Intensity–modulated Radiation Therapy with Pelvis Mask. J Sakon Nak Hosp [Internet]. 2024 Jun. 25 [cited 2024 Dec. 22];25(3):26-35. Available from: https://he05.tci-thaijo.org/index.php/JSakonNakHosp/article/view/2755

Issue

Section

Original article