Assessment of the Konsung hemoglobin analyzer’s accuracy and precision compared to the standard method in measuring hemoglobin and hematocrit levels
Keywords:
Konsung Hemoglobin Analyzer, hemoglobin, hematocrit, accuracy, precision, POCTAbstract
Background: The measurement of hemoglobin (Hb) and hematocrit (Hct) is essential for the diagnosis and monitoring of anemia. Automated hematology analyzers and the microhematocrit centrifuge are considered standard methods. However, point-of-care testing (POCT) devices such as the Konsung hemoglobin analyzer have gained increasing interest due to their convenience and rapid turnaround time. This study aimed to evaluate the accuracy and precision of the Konsung hemoglobin analyzer compared with standard methods.
Methods: This quasi-experimental analytical study included 100 patient blood samples, classified into three groups based on hemoglobin levels: low (male < 13.0 g/dL, female < 12.0 g/dL; n = 46), normal (male 13.0–16.5 g/dL, female 12.0–15.0 g/dL; n = 34), and high (male > 16.5 g/dL, female > 15.0 g/dL; n = 20). Hemoglobin and hematocrit levels were measured using the Konsung hemoglobin analyzer and compared with the Sysmex XN-3000 and the microhematocrit centrifuge within 2 hours after sample collection. Correlation analysis was performed using Pearson’s correlation coefficient. Agreement was evaluated using Bland–Altman analysis, including bias and limits of agreement (LoA). Precision was assessed using the coefficient of variation (%CV).
Results: Hematocrit values showed a strong correlation with the standard method (r = 0.975, p < 0.001), while hemoglobin values demonstrated a very strong correlation (r = 0.985, p < 0.001). Hematocrit showed a slight positive bias compared with the microhematocrit centrifuge method (+0.36%) and the Sysmex XN-3000 (+0.19%). The limits of agreement (LoA) for hematocrit ranged from -3.26% to +3.98% compared with the microhematocrit centrifuge method, and from -3.44% to +3.82% compared with the XN-3000. Hemoglobin measurements showed a positive bias of +0.49 g/dL compared with the XN-3000, with LoA ranging from -0.49 to +1.47 g/dL, all within clinically acceptable limits. Precision analysis demonstrated %CV values of less than 7% at all levels.
Conclusion: The Konsung hemoglobin analyzer demonstrated acceptable accuracy and precision compared with standard methods and can be used as a point-of-care testing device to support rapid clinical assessment. However, confirmatory testing using standard methods should be considered in cases with significant clinical implications.
References
Rituraj, Kumar AK, Singh PK. A comparative evaluation of various techniques for haemoglobin estimation. Int J Pharm Qual Assur 2025;16(2):387–392.
Sanchis-Gomar F, Cortell-Ballester J, Pareja-Galeano H, Banfi G, Lippi G. Hemoglobin point-of-care testing: the HemoCue system. J Lab Autom 2013;18(3):198–205.
Arhamidolatabadi T, Shams Vahdati S, Rikhtegar R. Point-of-care hemoglobin measurement in comparison with hematology analyzer: a cross-sectional study in emergency department. J Fam Med Prim Care 2019;8(9):2902–2906.
Dolscheid-Pommerich RC, Klauke S, Stoffel-Wagner B, et al. Comparability of point-of-care versus central laboratory hemoglobin determination in emergency patients at a supra-maximal care hospital. PLoS One 2016;11(2):e0149129.
Liu X, Wang S, Zhang Y, et al. Validation of a portable hemoglobin photometer (Konsung) for use in field conditions. Lancet Reg Health West Pac 2023;100789.
Morris SS, Ruel MT, Cohen RJ, Dewey KG, De la Brière B, Hassan MN. Precision, accuracy, and reliability of hemoglobin assessment with use of capillary blood. Am J Clin Nutr 1999;69(6):1243–1248.
Sari M, de Pee S, Martini E, Herman S, Sugiatmi, Bloem MW, et al. Estimating the prevalence of anaemia: a comparison of three methods. Bull World Health Organ 2001;79(6):506–511.
Sheth TN, Choudhry NK, Bowes M, Detsky AS. The relation of conjunctival pallor to the presence of anemia. J Gen Intern Med 1997;12(2):102–106.
Kumar N, Kushwaha R, Verma S, Singh S. Comparative evaluation of hemoglobin estimation and hematocrit determination using automated and manual methods. Clin Epidemiol Glob Health 2017;5(4):152–156.
Ahmad NA, Awaluddin SM, Samad R, Kasim NM, Yusof M, Aznuddin Abd Razak M, et al. Validity of point-of-care testing Mission Plus Hb in detecting anemia. Int J Biomedicine 2015;5(2):91–94.
Bhaskaram P, Moorthy V, Singh R, et al. Evaluation of the accuracy and reliability of hemoglobin measurement using the HemoCue system in a community setting. Lancet Reg Health West Pac 2020;4:100046.
Downloads
Published
How to Cite
Issue
Section
License
บทความที่ตีพิมพ์ถือว่าเป็นลิขสิทธิ์ของวารสารการแพทย์โรงพยาบาลสกลนคร การคัดลอกเพื่อพัฒนาเชิงวิชาการต้องได้รับการอ้างอิงอย่างถูกต้อง